DocumentCode :
544941
Title :
Measurement of potentials in the implanted ear: Models and pitfalls
Author :
Spelman, Francis A. ; Lamping, Jeffrey ; Rodenhiser, Kristin L.
Author_Institution :
Regional Primate Res. Center, Univ. of Washington, Seattle, WA, USA
Volume :
4
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
1312
Lastpage :
1313
Abstract :
Focused electrical stimuli will likely improve the performance of cochlear implants. Models predict the probable success of focusing. The models must be tested by measuring potentials in the inner ear. Successful testing requires the minimization of electrode impedances, the maximization of the common mode rejection ratios of the amplifiers used and the consistent selection of volt- age reference points in the subjects.
Keywords :
bioelectric potentials; cochlear implants; cochlear implants; focused electrical stimuli; implanted ea; inner ear; potentials measurement; voltage reference points; Electrodes;