DocumentCode
545282
Title
FM Doppler ultrasound system
Author
Pedersen, P.C. ; Wilhjelm, J.E.
Author_Institution
Dept. of Electrical Engineering, Worcester Polytechnic Institute, Worcester, MA 01609
Volume
5
fYear
1992
fDate
Oct. 29 1992-Nov. 1 1992
Firstpage
2120
Lastpage
2121
Abstract
This paper describes an ultrasound Doppler system utilizing swept frequency (FM) excitation. Both range and velocity data can be obtained from the received signals by means of either of two different signal processing approaches: Frequency (or spectral) Shift Measurements (FM-fsm) or Phase Shift Measurements (FM-psm). The unique features of the FM Doppler system are the low peak power, possibility for pre-compensation of attenuation and transducer effects, and — for the FM-fsm processing technique — the absence of aliasing effects.
Keywords
Doppler effect; Echo cancellers; Frequency modulation; Signal resolution; Ultrasonic imaging; Weight measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location
Paris, France
Print_ISBN
0-7803-0785-2
Electronic_ISBN
0-7803-0816-6
Type
conf
DOI
10.1109/IEMBS.1992.5762196
Filename
5762196
Link To Document