• DocumentCode
    545339
  • Title

    A new high-performance scan conversion method

  • Author

    Beni, Y. ; Abe, Y. ; Murakami, K. ; Shimura, T.

  • Author_Institution
    Medical System Laboratory, Fujitsu Laboratories Ltd. 1015 Kamikodanaka, Nakahara-ku, kawasaki, 211, Japan
  • Volume
    5
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    2249
  • Lastpage
    2250
  • Abstract
    The high-performance scan conversion we developed provides superior image quality even for low ultrasound scanning line density. TV pixel data is generated by linear and third — order interpolation, linear interpolation (bi — linear interpolation) having been employed conventionally. The result of experiments showed that this method reduces ridge artifacts produced along the scan line in conventional bi — linear interpolation. Our improvement in the third-order interpolation algorithm enabled the circuit size to be reduced to a practical level for real-time equipment applications.
  • Keywords
    Biomedical imaging; Hardware; Spline;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5762259
  • Filename
    5762259