DocumentCode
561852
Title
Assessment of ECG quality on an Android platform
Author
Johannesen, Lars
Author_Institution
Center for Drug Evaluation & Res., FDA, Silver Spring, MD, USA
fYear
2011
fDate
18-21 Sept. 2011
Firstpage
433
Lastpage
436
Abstract
An algorithm to determine quality of ECGs can enable inexperienced nurses and paramedics to record ECGs of sufficient diagnostic quality. We propose an algorithm that can assess the quality of an ECG designed for an Android-based platform. The algorithm is based on previously established ECG quality metrics for quantifying ECG quality but designed in a way to make it efficient to run on a mobile platform. Using the training data set the proposed algorithm obtained a sensitivity of 91% and a specificity of 85%. Testing against the test data sets, resulted in a score of 0.88 (events 1 and 2) and 0.79 (events 3). The proposed algorithm discriminates between ECGs of good and bad quality, which could help diagnose patients earlier and reduce associated treatment costs.
Keywords
electrocardiography; medical computing; patient diagnosis; Android platform; ECG quality assessment; ECG quality metrics; ECG quality quantification; associated treatment costs; diagnostic quality; mobile platform; paramedics; patient diagnosis; test data sets; Cardiovascular diseases; Electrocardiography; Lead; Low-frequency noise; Noise measurement; Sensitivity;
fLanguage
English
Publisher
ieee
Conference_Titel
Computing in Cardiology, 2011
Conference_Location
Hangzhou
ISSN
0276-6547
Print_ISBN
978-1-4577-0612-7
Type
conf
Filename
6164595
Link To Document