    Performance of VF detection parameters in an algorithm design scenario and in a real resuscitation scenario

  • Author

    Ayala, Unai ; Irusta, Unai ; Alonso, Erik ; Gonzalez, Digna

  • Author_Institution
    Univ. of the Basque Country, Bilbao, Spain
  • fYear
    2011
  • fDate
    18-21 Sept. 2011
  • Firstpage
    545
  • Lastpage
    548
  • Abstract
    Detection of Ventricular fibrillation (VF) in automated external defibrillators (AED) is tested following the recommendations of the American Heart Association (AHA). However, nonshockable out-of-hospital cardiac arrest (OHCA) rhythms may be very different from those covered in the AHA recommendations. In this study we compare the performance of four VF detection parameters for a testing database of rhythms covered by the AHA recommendations and a database of OHCA rhythms. Spectral parameters performed better than parameters related to the heart rate or the complexity for the testing database but worse for the OHCA database. The performance of the parameters was very different in an algorithm design scenario compliant with the AHA statement and in a real resuscitation scenario.
  • Keywords
    cardiology; computational complexity; defibrillators; medical computing; medical signal processing; signal detection; American Heart Association; VF detection parameters; algorithm design scenario; automated external defibrillators; nonshockable out-of-hospital cardiac arrest rhythms; real resuscitation scenario; testing database complexity; ventricular fibrillation; Algorithm design and analysis; Complexity theory; Databases; Electrocardiography; Heart beat; Sensitivity; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computing in Cardiology, 2011
  • Conference_Location
    Hangzhou
  • ISSN
    0276-6547
  • Print_ISBN
    978-1-4577-0612-7
  • Type

    conf

  • Filename
    6164623