Abstract :
In this paper we present the more recent developments of a vapor cell frequency standard based on a pulsed technique. The laboratory prototype we implemented has shown very interesting metrological features, such as negligible light-shift and strongly reduced cavity-pulling. Moreover, a frequency stability of σy(τ) = 1.2×10-12 τ-1/2 for measurement times up to τ ≈ 100000 s has been achieved.
Keywords :
frequency stability; frequency standards; optical pumping; rubidium; time measurement; Rb; cavity-pulling reduction; frequency stability; metrological featur; pulsed optically pumped rubidium frequency standard; pulsed technique; time measurement; vapor cell frequency standard; Atom optics; Atomic clocks; Masers; Optical pumping; Pump lasers;