Title :
Non-contact respiration measurement using structured light 3-D sensor
Author :
Aoki, Hirooki ; Miyazaki, Masaki ; Nakamura, Hidetoshi ; Furukawa, Ryo ; Sagawa, Ryusuke ; Kawasaki, Hiroshi
Author_Institution :
Grad. Sch. of Inf. Sci., Hiroshima City Univ., Hiroshima, Japan
Abstract :
We have examined a non-contact respiration measurement using Kinect, that is 3D sensor based on structured light, in order to decrease the burden of the pulmonary function test. In our proposed method, the respiratory waveform is produced from the volume change of thoracoabdominal part of examinee measured by the three-dimensional measurement function of Kinect. In this paper, we examine the validity of our proposed method by simultaneous measurement with the expiration gas analyzer and the flow meter. The experimental results implies that quantitative respiration measurement without contact-type sensor is realized by Kinect.
Keywords :
biomedical measurement; chemical analysis; flow sensors; flowmeters; pneumodynamics; Kinect; expiration gas analyzer; flow meter; noncontact respiration measurement; pulmonary function test; structured light 3-D sensor; thoracoabdominal domain; three-dimensional measurement function; Biomedical imaging; Educational institutions; Fluid flow measurement; Hip; Shape; Thorax; Volume measurement; 3-D measurement; Respiration measurement; non-contact physiological measurement; structured light;
Conference_Titel :
SICE Annual Conference (SICE), 2012 Proceedings of
Conference_Location :
Akita
Print_ISBN :
978-1-4673-2259-1