DocumentCode :
580556
Title :
Pilot study of floor-reactive-force generator mounted on MRI compatible lower-extremity motion simulator
Author :
Ikeda, Takashi ; Matsushita, Akira ; Saotome, Kousaku ; Hasegawa, Yohei ; Sankai, Yoshiyuki
Author_Institution :
Dept. of Intell. Interaction Technol., Univ. of Tsukuba, Tsukuba, Japan
fYear :
2012
fDate :
7-12 Oct. 2012
Firstpage :
311
Lastpage :
316
Abstract :
This paper describes a novel motion simulator for the lower extremities of human in a magnetic resonance imaging (MRI) environment. This motion simulator provides a wearer with physical supports to move their lower extremities or physical constrains as well as floor reactive force on bottoms of their feet during gait-like motion in an MRI room so that brain activities could be simultaneously measured. An MRI is one of the most powerful tools to measure activities in any part of brain but a device attached on a subject is limited because the material used for the device should be nonmagnetic. This paper shows the compatibility of the motion simulator that consists of Mckibben-type pneumatic artificial muscles and nonmagnetic materials. Also this paper shows the performance of the floor-reactive-force generator mounted on the soles of the motion simulator.
Keywords :
biomedical MRI; brain; electroactive polymer actuators; gait analysis; patient monitoring; MRI compatible lower extremity motion simulator; Mckibben type pneumatic artificial muscle; floor reactive force generator; gait like motion; magnetic resonance imaging environment; nonmagnetic material; Brain; Force; Generators; Hip; Joints; Magnetic resonance imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Intelligent Robots and Systems (IROS), 2012 IEEE/RSJ International Conference on
Conference_Location :
Vilamoura
ISSN :
2153-0858
Print_ISBN :
978-1-4673-1737-5
Type :
conf
DOI :
10.1109/IROS.2012.6385626
Filename :
6385626
Link To Document :
بازگشت