DocumentCode :
584158
Title :
Whole-body averaged SAR measurements for small phantom by calorimetric method
Author :
Akiyama, Takuya ; Wake, Kanako ; Arima, Takuji ; Watanabe, Soichi ; Uno, Toru
Author_Institution :
Tokyo Univ. of Agric. & Technol., Koganei, Japan
fYear :
2012
fDate :
Oct. 29 2012-Nov. 2 2012
Firstpage :
692
Lastpage :
695
Abstract :
This paper has investigated on heat transfer medium for whole-body averaged SAR measurements by a calorimetric method. The microwave exposure experiment and measured WBA SAR (specific absorption rate) are indicated. It turned out that measurement accuracy and measuring time change by changing heat transfer medium. Furthermore, when using water as heat transfer medium, error with measured values and simulated values of WBA SAR are less than 10% and this is the best accuracy in used three kinds of heat transfer medium.
Keywords :
calorimetry; heat transfer; microwave measurement; calorimetric method; heat transfer medium; microwave exposure experiment; phantom; whole-body averaged specific absorption rate measurement; Heat transfer; Measurement uncertainty; Microwave measurements; Temperature measurement; Time measurement; Voltage measurement; Water heating;