• DocumentCode
    584158
  • Title

    Whole-body averaged SAR measurements for small phantom by calorimetric method

  • Author

    Akiyama, Takuya ; Wake, Kanako ; Arima, Takuji ; Watanabe, Soichi ; Uno, Toru

  • Author_Institution
    Tokyo Univ. of Agric. & Technol., Koganei, Japan
  • fYear
    2012
  • fDate
    Oct. 29 2012-Nov. 2 2012
  • Firstpage
    692
  • Lastpage
    695
  • Abstract
    This paper has investigated on heat transfer medium for whole-body averaged SAR measurements by a calorimetric method. The microwave exposure experiment and measured WBA SAR (specific absorption rate) are indicated. It turned out that measurement accuracy and measuring time change by changing heat transfer medium. Furthermore, when using water as heat transfer medium, error with measured values and simulated values of WBA SAR are less than 10% and this is the best accuracy in used three kinds of heat transfer medium.
  • Keywords
    calorimetry; heat transfer; microwave measurement; calorimetric method; heat transfer medium; microwave exposure experiment; phantom; whole-body averaged specific absorption rate measurement; Heat transfer; Measurement uncertainty; Microwave measurements; Temperature measurement; Time measurement; Voltage measurement; Water heating;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Antennas and Propagation (ISAP), 2012 International Symposium on
  • Conference_Location
    Nagoys
  • Print_ISBN
    978-1-4673-1001-7
  • Type

    conf

  • Filename
    6394033