DocumentCode :
585302
Title :
EMF dose in patients and medical staff during hyperthermia treatment of cancer
Author :
Bakker, Jurriaan ; Canters, Richard ; Neufeld, Esra ; Paulides, Maarten ; Van Rhoon, Gerard
Author_Institution :
Dept. of Radiat. Oncology, Erasmus MC - Daniel den Hoed Cancer Center, Rotterdam, Netherlands
fYear :
2012
fDate :
17-21 Sept. 2012
Firstpage :
1
Lastpage :
5
Abstract :
At the Erasmus Medical Center, we apply hyperthermia treatments (HT) of cancer in the pelvic area using an electromagnetic applicator installed in a Faraday shielded treatment room. Consequently, medical staff and accompanying persons are exposed to electromagnetic fields (EMF) during treatment. In the past, procedures were defined based on measurements defining compliant areas where the fields are expected to be below the exposure guidelines. Today, advanced electromagnetic models enable more precise dose assessment in the human anatomy and better evaluation of the procedures. The objective of this study was to investigate theoretically if the exposure of medical staff and accompanying persons is compliant to the guidelines when adhering to the procedures. Therefore, the whole-body averaged (SARwb) and spatial-averaged (SAR10g) specific absorption rate was assessed numerically in whole-body models of the patient, medical staff and an accompanying person. We found that predicted fields can exceed the exposure guidelines ten fold. However, the basic restrictions on SARwb and SAR10g are not exceeded for workers, but SARwb can exceed the guideline by a factor of two for accompanying persons that are in close vicinity of the applicator. In conclusion, adhering to the procedures defined in the early nineties is effective in keeping the exposure in medical staff and accompanying persons compliant to the exposure guidelines.
Keywords :
cancer; dosimetry; electromagnetic fields; finite difference time-domain analysis; hyperthermia; occupational health; radiation therapy; EMF dose; Faraday shielded treatment; advanced electromagnetic models; cancer; compliant areas; dose assessment; electromagnetic applicator; electromagnetic fields; human anatomy; hyperthermia treatment; medical staff; pelvic area; spatial-averaged specific absorption rate; whole-body averaged specific absorption rate; whole-body models; Applicators; Atmospheric modeling; Electromagnetics; Guidelines; Head; Hyperthermia; Planning;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetic Compatibility (EMC EUROPE), 2012 International Symposium on
Conference_Location :
Rome
ISSN :
2325-0356
Print_ISBN :
978-1-4673-0718-5
Type :
conf
DOI :
10.1109/EMCEurope.2012.6396793
Filename :
6396793
Link To Document :
بازگشت