In Vitro and in Vivo tissue harmonic images obtained with parallel transmit beamforming by means of orthogonal frequency division multiplexing [Correspondence]
Author :
Demi, Libertario ; Ramalli, Alessandro ; Giannini, Gabriele ; Mischi, Massimo
Author_Institution :
Lab. of Biomed. Diagnostics, Eindhoven Univ. of Technol., Eindhoven, Netherlands
Volume :
62
Issue :
1
fYear :
2015
fDate :
Jan-15
Firstpage :
230
Lastpage :
235
Abstract :
In classic pulse-echo ultrasound imaging, the data acquisition rate is limited by the speed of sound. To overcome this, parallel beamforming techniques in transmit (PBT) and in receive (PBR) mode have been proposed. In particular, PBT techniques, based on the transmission of focused beams, are more suitable for harmonic imaging because they are capable of generating stronger harmonics. Recently, orthogonal frequency division multiplexing (OFDM) has been investigated as a means to obtain parallel beamformed tissue harmonic images. To date, only numerical studies and experiments in water have been performed, hence neglecting the effect of frequencydependent absorption. Here we present the first in vitro and in vivo tissue harmonic images obtained with PBT by means of OFDM, and we compare the results with classic B-mode tissue harmonic imaging. The resulting contrast-to-noise ratio, here used as a performance metric, is comparable. A reduction by 2 dB is observed for the case in which three parallel lines are reconstructed. In conclusion, the applicability of this technique to ultrasonography as a means to improve the data acquisition rate is confirmed.
Keywords :
OFDM modulation; array signal processing; biological tissues; biomedical ultrasonics; B-mode tissue harmonic imaging; contrast-to-noise ratio; data acquisition rate; focused beam transmission; frequency-dependent absorption effect; orthogonal frequency division multiplexing; parallel beamformed tissue harmonic images; parallel lines; parallel transmit beamforming; performance metric; pulse-echo ultrasound imaging; receive mode; sound speed; transmit mode; ultrasonography; water; Array signal processing; Cavity resonators; Harmonic analysis; Imaging; OFDM; Power harmonic filters; Ultrasonic imaging;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2014.006599
Filename :
7002940
Link To Document :
بازگشت