Title : 
Improvement of magnetic force microscope resolution and application to high-density recording media
         
        
            Author : 
Futamoto, Masaaki ; Hagami, T. ; Ishihara, Sayaka ; Ohtake, T.
         
        
            Author_Institution : 
Fac. of Sci. & Eng., Chuo Univ., Tokyo, Japan
         
        
        
            fDate : 
Oct. 31 2012-Nov. 2 2012
         
        
        
        
            Abstract : 
Effects of magnetic material, coating thickness, and tip radius on magnetic force microscope (MFM) resolution have been investigated. MFM resolutions better than 8 nm have been confirmed with tips coated by high-magnetic moment materials with optimized thicknesses. High-resolution tips are applied to the observations of recorded magnetization structures of perpendicular and bit-patterned media samples.
         
        
            Keywords : 
magnetic force microscopy; magnetic materials; magnetic moments; bitpatterned media samples; coating thickness; high-density recording media; high-magnetic moment materials; high-resolution tips; magnetic force microscope resolution; magnetic material effect; magnetization structures; tip radius; Facsimile; Magnetic force microscopy; Magnetic resonance imaging; Microscopy; Spatial resolution; Switches; magnetic force microscope; perpendicular recording media; spatial resolution; switching field;
         
        
        
        
            Conference_Titel : 
APMRC, 2012 Digest
         
        
            Conference_Location : 
Singapore
         
        
            Print_ISBN : 
978-1-4673-4734-1