  • Abstract
    In point-based registration between the patient and the image, skin -affixed fiducial markers can only be attached to the patient skin. Registration accuracy is relatively high in the region around the markers. However, the accuracy gradually decreases proportionally to the distance from the surface where the markers are attached. To decrease the target registration error (TRE) rather than the fiducial registration error (FRE) around the target, reference points should be located near the target organ. In order to decrease the TRE, a method is proposed that makes it possible to register an anatomical landmark as a reference point around the target using ultrasonography.
  • Keywords
    biomedical ultrasonics; image registration; medical image processing; patient care; skin; FRE; TRE; anatomical landmark; fiducial registration error; high accuracy patient image registration; patient skin; point based registration; registration accuracy; skin affixed fiducial markers; target registration error; ultrasonography; Accuracy; Adaptive optics; Calibration; Optical imaging; Phantoms; Probes; Ultrasonic imaging; High accuracy registration; Registration; Surgical navigation; Ultrasound registration;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ubiquitous Robots and Ambient Intelligence (URAI), 2012 9th International Conference on
  • Conference_Location
    Daejeon
  • Print_ISBN
    978-1-4673-3111-1
  • Electronic_ISBN
    978-1-4673-3110-4
  • Type

    conf

  • DOI
    10.1109/URAI.2012.6462934
  • Filename
    6462934