• DocumentCode
    598333
  • Title

    A radiation detection readout circuit with current feedback baseline holder

  • Author

    Xiao-Lu Chen ; Ya-Cong Zhang ; Wen-Gao Lu ; Zhong-Jian Chen

  • Author_Institution
    Key Lab. of Microelectron. Devices & Circuits, Peking Univ., Beijing, China
  • fYear
    2012
  • fDate
    Oct. 29 2012-Nov. 1 2012
  • Firstpage
    1
  • Lastpage
    3
  • Abstract
    This paper describes a radiation detection readout circuit for portable dosimeter which is aimed at low power, low noise and high counting rate. A current feedback baseline holder circuit is proposed to solve the baseline shift problem without any other performance deterioration. The core circuit has been implemented in 0.35μm CMOS technology. It achieves 46mV/fC conversion gain, 200kcps counting rate and consumes 260μA current from 3.3V supply. When no detector is connected to the chip, the equivalent input noise charge is 0.094fC rms.
  • Keywords
    CMOS integrated circuits; dosimeters; radiation detection; readout electronics; CMOS technology; baseline shift problem; core circuit; current 260 muA; current feedback baseline holder; high counting rate; low noise rate; low power rate; portable dosimeter; radiation detection readout circuit; size 0.35 mum; voltage 3.3 V; CMOS integrated circuits; CMOS technology; Detectors; Power demand; Signal to noise ratio; Topology;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State and Integrated Circuit Technology (ICSICT), 2012 IEEE 11th International Conference on
  • Conference_Location
    Xi´an
  • Print_ISBN
    978-1-4673-2474-8
  • Type

    conf

  • DOI
    10.1109/ICSICT.2012.6467691
  • Filename
    6467691