The feasibility of application phonoangiography with fractional order chaotic system in arteriovenous shunt stenosis detection
Author :
Wei-Ling Chen ; Chung-Dann Kan ; Chia-Hung Lin ; Tainsong Chen
Author_Institution :
Dept. of Biomed. Eng., Nat. Cheng Kung Univ., Tainan, Taiwan
fYear :
2013
fDate :
12-16 March 2013
Firstpage :
123
Lastpage :
126
Abstract :
This paper proposed the signal processing method for evaluation of arteriovenous shunt (AVS) stenosis in hemodialysis patients. AV shunts are surgically created pathological or physiological fistulas serving as access routes for end-stage renal disease (ESRD) patients. The distinct and periodic bruit of vascular shunts is clearly audible over the access routes. Thus, a bruit spectral analysis could be considered a valuable noninvasive method for quantifying the severity of vessel stenosis. This study comprised the Burg method and fractional order chaos system (FOCS). We utilized the Burg method to estimate the characteristic frequency of bruit and a FOCS to monitor the differing frequency spectra between the normal condition and AV access stenosis. This study recruited 50 years old female and received the regular follow-up analysis for her forearm loop AVS function during 5 months. From these results, we observed that the function of AVS can be estimated by the butterfly patterns. This received the regular follow-up analysis for her forearm loop AVS function.
Keywords :
diseases; medical signal detection; medical signal processing; spectral analysis; surgery; Burg method; ESRD patients; FOCS; arteriovenous shunt stenosis detection; end-stage renal disease patients; forearm loop AVS function; fractional order chaos system; fractional order chaotic system; hemodialysis patients; pathological fistulas; periodic bruit spectral analysis; phonoangiography; physiological fistulas; signal processing; surgery; time 5 month; valuable noninvasive method; vascular shunts; vessel stenosis; Chaos; Cities and towns; Educational institutions; Frequency estimation; Lesions; Mathematical model; Stethoscope; AVS Stenosis; Arteriovenous Shunt (AVS); Burg Method; Percutaneous Transluminal Angioplasty (PTA); Phonoangiographic Signal;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Orange Technologies (ICOT), 2013 International Conference on
Conference_Location :
Tainan
Print_ISBN :
978-1-4673-5934-4
Type :
conf
DOI :
10.1109/ICOT.2013.6521173
Filename :
6521173
Link To Document :
بازگشت