• DocumentCode
    604158
  • Title

    Design of a Portable Wireless EEG System Using a Fully Integrated Analog Front End

  • Author

    Davies, P.J. ; Bohorquez, Jorge

  • Author_Institution
    Dept. of Biomed. Eng., Univ. of Miami, Miami, FL, USA
  • fYear
    2013
  • fDate
    3-5 May 2013
  • Firstpage
    63
  • Lastpage
    64
  • Abstract
    This paper describes the design of a wireless Electroencephalograph (EEG) system using a new generation of high resolution Analog Front End circuits that minimize signal acquisition hardware and are aimed at simplifying recordings from patients by not requiring them to be tied down with current bulky recording hardware. Such a modality opens up EEG monitoring to more ambulatory themed studies, and further solidifies the future of EEG for diagnostic applications by providing a way for investigators to easily record from active subjects and be less intrusive while doing so. Design was validated with Alpha Band monitoring when eyes are open vs. closed and observing a 40dB increases in 12Hz RMS power when eyes are closed.
  • Keywords
    electroencephalography; medical signal detection; Alpha Band monitoring; electroencephalograph; frequency 12 kHz; fully integrated analog front end; noise figure 40 dB; portable wireless EEG system design; signal acquisition hardware; Batteries; Electrodes; Electroencephalography; Microcontrollers; Monitoring; Signal resolution; Wireless communication;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering Conference (SBEC), 2013 29th Southern
  • Conference_Location
    Miami, FL
  • Print_ISBN
    978-1-4799-0624-6
  • Type

    conf

  • DOI
    10.1109/SBEC.2013.40
  • Filename
    6525677