Title :
Comparison of magnetostimulation limits for axial and transverse drive fields in MPI
Author :
Yu, Elaine ; Saritas, E.U. ; Conolly, S.M.
Author_Institution :
Dept. of Bioeng., Univ. of California, Berkeley, Berkeley, CA, USA
Abstract :
As shown in Fig. 1b, the mean magneto stimulation thresholds for the transverse field was approximately 3 times lower than the axial field (ΔBmin = 26 mT-pp vs. 78 mT-pp) for the same set of subjects, with comparable τc values (265 μs vs. 243 μs). In addition, the percentage deviation across subjects was significantly smaller for the transverse field (13%) than the axial field (28%). For both transverse and axial fields, the data trends were well modeled by the hyperbolic amplitude vs. frequency curve, as demonstrated in Fig. 1c for a single subject.
Keywords :
biomagnetism; biomedical imaging; MPI; axial drive field; magneto stimulation limit; transverse drive field; Biomedical engineering; Coils; Educational institutions; Magnetic field measurement; Magnetic flux; Magnetic resonance imaging;
Conference_Titel :