DocumentCode
605670
Title
Approximated elliptical coils in magnetic particle imaging
Author
Kaethner, Christian ; Grafe, K. ; Gruttner, M. ; Buzug, Thorsten M.
Author_Institution
Inst. of Med. Eng., Univ. of Lubeck, Lubeck, Germany
fYear
2013
fDate
23-24 March 2013
Firstpage
1
Lastpage
1
Abstract
The paper presented results of simulated and measured magnetic fields of an approximated elliptical coil. The simulation values are coincided with the measurement values. Using the relative error as difference metric, the difference between simulation and measurement is presented as well. The visual comparison between the images as well as the resulting values of the relative error show only a minimal difference in the inner zone of the coils. The gradient values for the prototype coil arrangement for x, y and z direction were also shown. With an increasing side length the gradients in x and y direction decrease, while the gradient in z direction slightly increases.
Keywords
coils; image registration; magnetic fields; medical image processing; approximated elliptical coils; gradient values; magnetic field measurement; magnetic particle imaging; prototype coil arrangement; Approximation methods; Coils; Imaging; Magnetic field measurement; Magnetic fields; Magnetic particles; Measurement uncertainty;
fLanguage
English
Publisher
ieee
Conference_Titel
Magnetic Particle Imaging (IWMPI), 2013 International Workshop on
Conference_Location
Berkeley, CA
Print_ISBN
978-1-4673-5520-9
Electronic_ISBN
978-1-4673-5521-6
Type
conf
DOI
10.1109/IWMPI.2013.6528343
Filename
6528343
Link To Document