First hybrid MPI-MRI imaging system as integrated design for mice and rats: Description of the instrumentation setup
Author :
Franke, J. ; Heinen, U. ; Matthies, L. ; Niemann, V. ; Jaspard, F. ; Heidenreich, M. ; Buzug, T.
Author_Institution :
Bruker BioSpin MRI GmbH, Ettlingen, Germany
fYear :
2013
fDate :
23-24 March 2013
Firstpage :
1
Lastpage :
1
Abstract :
This hybrid MPI-MRI imaging system (see Fig. 1 RIGHT) is currently under assembly at the research labs of Bruker BioSpin MRI GmbH, whereas single components were already tested. The magnetic field strength and the magnetic field gradient were measured to be up to Bz=0.495 T and up to Gz=2.2 T/m for the PF and the SF, respectively (see Fig. 1 LEFT). The MRI gradient strengths were determined to be Gx=0.84 mT/A/m, Gy=0.9 mT/A/m, Gz=0.87 mT/A/m (Imax=300 A) and the MRI RF coil was tuned to the Larmor frequency of protons at 0.495 T. The DF resonant circuits were tuned to excitation frequencies of around 25 kHz. With expected maximal DF amplitude of 20 mT and the strongest SF gradient strength a field of view of 18×36×36 mm3 could be reached.
Keywords :
biomedical MRI; biomedical electronics; biomedical equipment; coils; magnetic particles; Bruker BioSpin MRI GmbH research labs; DF resonant circuits; Larmor frequency; MRI RF coil; assembly; excitation frequencies; expected maximal DF amplitude; first hybrid MPI-MRI imaging system; instrumentation setup; integrated design; magnetic field gradient; magnetic field strength; magnetic flux density 0.495 T; magnetic flux density 20 mT; mice; rats; Biomedical imaging; Coils; Magnetic field measurement; Magnetic fields; Magnetic resonance imaging; Mice;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Magnetic Particle Imaging (IWMPI), 2013 International Workshop on
Conference_Location :
Berkeley, CA
Print_ISBN :
978-1-4673-5520-9
Electronic_ISBN :
978-1-4673-5521-6
Type :
conf
DOI :
10.1109/IWMPI.2013.6528363
Filename :
6528363
Link To Document :
بازگشت