DocumentCode
609805
Title
Thermosonic bonding for ultrasound transducers: Low-temperature metallurgical bonding
Author
Aasmundtveit, Knut E. ; Luu, Thi Thuy ; Eggen, Trym ; Baumgartner, Charles E ; Hoivik, Nils ; Wang, Kaiying ; Nguyen, Hoang-Vu ; Imenes, Kristin
Author_Institution
HiVe - Vestfold University College, IMST - Dept. of Micro and Nano Systems Technology, 3184 Borre, Norway
fYear
2012
fDate
17-20 Sept. 2012
Firstpage
1
Lastpage
6
Abstract
A low-temperature bonding process for ultrasound transducers is presented: compatible with poling requirements, manufacturability and reliability. In this work, we demonstrate that a thermosonic bonding process can provide a reliable, metallurgical bond at moderate temperatures, even down to room temperature, with bonding times in the order of seconds. Bonding parameters (temperature, compression force, ultrasonic energy) were optimized by evaluating shear strength on Au stud bump bonded Si chips. Model systems have been bonded, mimicking a complete Electro-Acoustic Module (EAM), including a stack of IC emulator / flex interconnection / interface part of the ultrasound transducer.
fLanguage
English
Publisher
ieee
Conference_Titel
Electronic System-Integration Technology Conference (ESTC), 2012 4th
Conference_Location
Amsterdam, Netherlands
Print_ISBN
978-1-4673-4645-0
Type
conf
DOI
10.1109/ESTC.2012.6542142
Filename
6542142
Link To Document