  • Title

    Automated transmission electron microscopy for defect review and metrology of Si devices

  • Author

    Strauss, Michael ; Williamson, Mark

  • Author_Institution
    FEI Co., Hillsboro, OR, USA
  • fYear
    2013
  • fDate
    14-16 May 2013
  • Firstpage
    366
  • Lastpage
    370
  • Abstract
    Functions to automate TEM sample alignment and focus have been developed and tested on a 200 kV S/TEM. Sensitivity and repeatability data for functions allowing for automated eucentric height, focus, and zone axis alignment are presented. Eucentric and focus autofunctions were used to collect dynamic precision metrology data from a commercially available 22nm microprocessor. Results show 3σ standard deviation dynamic precision values better than 0.5nm for all recorded measurements.
  • Keywords
    measurement; microprocessor chips; semiconductor industry; transmission electron microscopy; Si devices; automate TEM sample alignment; automated eucentric height; automated transmission electron microscopy; defect review; metrology; microprocessor; Logic gates; Metrology; Scanning electron microscopy; Semiconductor device measurement; Sensitivity; Standards; Automation; CD-TEM; Metrology; TEM;