DocumentCode
614924
Title
Automated transmission electron microscopy for defect review and metrology of Si devices
Author
Strauss, Michael ; Williamson, Mark
Author_Institution
FEI Co., Hillsboro, OR, USA
fYear
2013
fDate
14-16 May 2013
Firstpage
366
Lastpage
370
Abstract
Functions to automate TEM sample alignment and focus have been developed and tested on a 200 kV S/TEM. Sensitivity and repeatability data for functions allowing for automated eucentric height, focus, and zone axis alignment are presented. Eucentric and focus autofunctions were used to collect dynamic precision metrology data from a commercially available 22nm microprocessor. Results show 3σ standard deviation dynamic precision values better than 0.5nm for all recorded measurements.
Keywords
measurement; microprocessor chips; semiconductor industry; transmission electron microscopy; Si devices; automate TEM sample alignment; automated eucentric height; automated transmission electron microscopy; defect review; metrology; microprocessor; Logic gates; Metrology; Scanning electron microscopy; Semiconductor device measurement; Sensitivity; Standards; Automation; CD-TEM; Metrology; TEM;
fLanguage
English
Publisher
ieee
Conference_Titel
Advanced Semiconductor Manufacturing Conference (ASMC), 2013 24th Annual SEMI
Conference_Location
Saratoga Springs, NY
ISSN
1078-8743
Print_ISBN
978-1-4673-5006-8
Type
conf
DOI
10.1109/ASMC.2013.6552761
Filename
6552761
Link To Document