Abstract :
Modern Magnetic Resonance Imaging (MRI) methods now allow the rapid acquisition of images with an excellent tissue contrast and high spatial resolution. Complex organ deformations can thus be estimated using image registration techniques applied to anatomical information. This opens great perspectives for the use of MRI to retroactively target an interventional procedure in mobile organs in real-time. For this purpose, both the update time and the latency of the motion information are two key points. In the current paper, the organ deformation is estimated on a voxel-by-voxel basis and a Kalman predictor is used to compensate for the residual latency. The implementation benefitted from the parallel architecture of Graphical Processing Units (GPU) for accelerating computation times. The efficiency and the potential of the method to anticipate organ displacements in real-time was evaluated on the abdomen of twelve free-breathing volunteers. The deformation of both kidney and liver could be updated with a rate of 10 Hz over sustained periods of several minutes, and the employed Kalman predictor reduced the tracking error in average by 30%.
Keywords :
Kalman filters; biological tissues; biomedical MRI; graphics processing units; image motion analysis; image registration; image resolution; kidney; liver; medical image processing; pneumodynamics; GPU; Kalman predictor; MRI; abdomen; anatomical information; complex organ deformations; free-breathing volunteers; graphical processing units; image registration; interventional procedures; kidney; liver; magnetic resonance imaging guidance; mobile organs; motion information; organ displacement; parallel architecture; rapid image acquisition; real-time anticipation; residual latency; spatial resolution; tissue contrast; tracking error; voxel-by-voxel basis; Biomedical imaging; Kalman filters; Kidney; Liver; Magnetic resonance imaging; Real-time systems; Tracking; Magnetic Resonance Imaging; Motion analysis; Real time systems;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Imaging (ISBI), 2013 IEEE 10th International Symposium on
Conference_Location :
San Francisco, CA
ISSN :
1945-7928
Print_ISBN :
978-1-4673-6456-0
Type :
conf
DOI :
10.1109/ISBI.2013.6556800
Filename :
6556800
Link To Document :
بازگشت