Title :
Miniaturization of package for an implantable heart monitoring device
Author :
Nguyen, Anh-Tuan T. ; Tjulkins, Fjodors ; Aasmundtveit, Knut E. ; Hoivik, Nils ; Hoff, Lars ; Imenes, Kristin
Author_Institution :
Dept. of Micro & Nano Syst. Technol., Oslo Univ. Hosp., Tonsberg, Norway
Abstract :
A very first myocardial (heart muscle) implantable acceleration prototype-device, using a commercial MEMS accelerometer is presented in this paper. Procedures of design, fabrication, assembly and testing of the prototype are included in detail. Preliminary results of tests based on the regulations for implantable devices have been done successfully. The paper concludes with directions for improving the device.
Keywords :
accelerometers; bioMEMS; biomedical electronics; cardiology; patient monitoring; prosthetics; prototypes; commercial MEMS accelerometer; device improvement; heart muscle implantable acceleration prototype-device; implantable device regulation; implantable heart monitoring device; myocardial implantable acceleration prototype-device; package miniaturization; prototype assembly procedure; prototype design procedure; prototype fabrication procedure; prototype testing procedure; test preliminary result; Accelerometers; Capacitors; Current measurement; Encapsulation; Heart; Leakage currents; Substrates; MEMS accelerometer; Myocardial; implantable acceleration device;
Conference_Titel :
Design, Test, Integration and Packaging of MEMS/MOEMS (DTIP), 2013 Symposium on
Conference_Location :
Barcelona
Print_ISBN :
978-1-4673-4477-7