• DocumentCode
    621509
  • Title

    Efficient evaluation of MRI-induced electric fields in the vicinity of implantable lead

  • Author

    Shi Feng ; Rui Qiang ; Kainz, Wolfgang ; Jackson, D. ; Ji Chen

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Univ. of Houston, Houston, TX, USA
  • fYear
    2013
  • fDate
    4-5 April 2013
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    This paper presents a novel technique for the efficient evaluation of MRI-induced electric fields in the vicinity of implantable metallic leads. The technique is based on the reciprocity theorem in conjuncture with the Huygens principle that allows one to separate the micro-scale modeling of the metallic lead from the macro-level modeling of human subjects within MRI scanners. In addition to the numerical efficiency, this approach clearly explains the lead heating mechanism during MRI operations. Numerical examples are used to demonstrate the efficiency and accuracy of this proposed approach.
  • Keywords
    bioelectric phenomena; biomedical MRI; numerical analysis; prosthetics; Huygens principle; MRI operations; MRI scanners; MRI-induced electric field evaluation; human subjects; implantable metallic leads; lead heating mechanism; macrolevel modeling; magnetic resonance imaging; microscale modeling; numerical efficiency; reciprocity theorem; Biomedical imaging; Coils; Computational modeling; Electric fields; Heating; Magnetic resonance imaging; Radio frequency; implantable lead; magnetic resonance imaging (MRI) safety; numerical modeling; reciprocity theorem;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Wireless and Microwave Circuits and Systems (WMCS), 2013 Texas Symposium on
  • Conference_Location
    Waco, TX
  • Print_ISBN
    978-1-4799-0456-3
  • Type

    conf

  • DOI
    10.1109/WMCaS.2013.6563554
  • Filename
    6563554