DocumentCode
621509
Title
Efficient evaluation of MRI-induced electric fields in the vicinity of implantable lead
Author
Shi Feng ; Rui Qiang ; Kainz, Wolfgang ; Jackson, D. ; Ji Chen
Author_Institution
Dept. of Electr. & Comput. Eng., Univ. of Houston, Houston, TX, USA
fYear
2013
fDate
4-5 April 2013
Firstpage
1
Lastpage
4
Abstract
This paper presents a novel technique for the efficient evaluation of MRI-induced electric fields in the vicinity of implantable metallic leads. The technique is based on the reciprocity theorem in conjuncture with the Huygens principle that allows one to separate the micro-scale modeling of the metallic lead from the macro-level modeling of human subjects within MRI scanners. In addition to the numerical efficiency, this approach clearly explains the lead heating mechanism during MRI operations. Numerical examples are used to demonstrate the efficiency and accuracy of this proposed approach.
Keywords
bioelectric phenomena; biomedical MRI; numerical analysis; prosthetics; Huygens principle; MRI operations; MRI scanners; MRI-induced electric field evaluation; human subjects; implantable metallic leads; lead heating mechanism; macrolevel modeling; magnetic resonance imaging; microscale modeling; numerical efficiency; reciprocity theorem; Biomedical imaging; Coils; Computational modeling; Electric fields; Heating; Magnetic resonance imaging; Radio frequency; implantable lead; magnetic resonance imaging (MRI) safety; numerical modeling; reciprocity theorem;
fLanguage
English
Publisher
ieee
Conference_Titel
Wireless and Microwave Circuits and Systems (WMCS), 2013 Texas Symposium on
Conference_Location
Waco, TX
Print_ISBN
978-1-4799-0456-3
Type
conf
DOI
10.1109/WMCaS.2013.6563554
Filename
6563554
Link To Document