Super-resolution imaging using multi- electrode CMUTs: theoretical design and simulation using point targets
Author :
Wei You ; Cretu, Edmond ; Rohling, Robert
Author_Institution :
Dept. of Electr. & Comput. Eng., Univ. of British Columbia, Vancouver, BC, Canada
Volume :
60
Issue :
11
fYear :
2013
fDate :
Nov-13
Firstpage :
2295
Lastpage :
2309
Abstract :
This paper investigates a low computational cost, super-resolution ultrasound imaging method that leverages the asymmetric vibration mode of CMUTs. Instead of focusing on the broadband received signal on the entire CMUT membrane, we utilize the differential signal received on the left and right part of the membrane obtained by a multi-electrode CMUT structure. The differential signal reflects the asymmetric vibration mode of the CMUT cell excited by the nonuniform acoustic pressure field impinging on the membrane, and has a resonant component in immersion. To improve the resolution, we propose an imaging method as follows: a set of manifold matrices of CMUT responses for multiple focal directions are constructed off-line with a grid of hypothetical point targets. During the subsequent imaging process, the array sequentially steers to multiple angles, and the amplitudes (weights) of all hypothetical targets at each angle are estimated in a maximum a posteriori (MAP) process with the manifold matrix corresponding to that angle. Then, the weight vector undergoes a directional pruning process to remove the false estimation at other angles caused by the side lobe energy. Ultrasound imaging simulation is performed on ring and linear arrays with a simulation program adapted with a multi-electrode CMUT structure capable of obtaining both average and differential received signals. Because the differential signals from all receiving channels form a more distinctive temporal pattern than the average signals, better MAP estimation results are expected than using the average signals. The imaging simulation shows that using differential signals alone or in combination with the average signals produces better lateral resolution than the traditional phased array or using the average signals alone. This study is an exploration into the potential benefits of asymmetric CMUT responses for super-resolution imaging.
Keywords :
capacitive sensors; electrodes; membranes; ultrasonic imaging; ultrasonic transducers; vibrational modes; asymmetric vibration mode; differential signal; directional pruning process; membrane; multielectrode CMUT; nonuniform acoustic pressure field; point targets; super-resolution ultrasound imaging method; theoretical design; weight vector; Acoustics; Arrays; Image resolution; Imaging; Manifolds; Signal resolution; Vectors; Computer Simulation; Image Processing, Computer-Assisted; Phantoms, Imaging; Signal Processing, Computer-Assisted; Transducers; Ultrasonography;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2013.6644734
Filename :
6644734
Link To Document :
بازگشت