DocumentCode :
627997
Title :
Low-Cost, Thermistor Based Respiration Monitor
Author :
Gupta, Madhu ; Qudsi, H.
Author_Institution :
Electr. & Comput. Eng., Cornell Univ., Ithaca, NY, USA
fYear :
2013
fDate :
5-7 April 2013
Firstpage :
287
Lastpage :
288
Abstract :
Respiration rate is a standard physiological measurement taken for monitoring a patient. Existing respiration monitors used in medical centers are too expensive to be used in low-resource environments. We have developed a low-cost respiratory monitor to address this issue. This device calculates a patient´s breathing rate by detecting changes in temperature through a thermistor when the patient breathes in a mask. Features of the device include an alarm through a piezoelectric speaker which sounds when the patient stops breathing and a low-battery indicator signal for when the battery powering the device dips below a threshold voltage. The design and implementation involved several different hardware components, as well as software functions. Our initial prototype of our project accomplished our main objectives, and experimental measurements indicate that the device measures a person´s respiration rate with relative accuracy.
Keywords :
alarm systems; biomedical equipment; biomedical measurement; biothermics; hardware-software codesign; microcontrollers; patient monitoring; physiology; piezoelectric devices; pneumodynamics; thermistors; alarm system; hardware component; low-battery indicator signal; low-cost respiratory monitor; low-resource environments; medical centers; patient breathing rate; patient monitoring; person respiration rate measurement; piezoelectric speaker; software function; standard physiological measurement; temperature change detection; thermistor based respiration monitor; threshold voltage; Batteries; Biomedical monitoring; Monitoring; Temperature measurement; Temperature sensors; Thermistors; Voltage measurement; biomedical; microcontroller; respiration; thermistor;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioengineering Conference (NEBEC), 2013 39th Annual Northeast
Conference_Location :
Syracuse, NY
ISSN :
2160-7001
Print_ISBN :
978-1-4673-4928-4
Type :
conf
DOI :
10.1109/NEBEC.2013.111
Filename :
6574471
Link To Document :
بازگشت