Four electrode embedded bioimpedance measurement system
Author
Margo, Cedric ; Katrib, Juliano ; Nadi, M. ; Rouane, Amar
Author_Institution
LPEM, ESPCI-ParisTech, Paris, France
fYear
2013
fDate
20-21 June 2013
Firstpage
1
Lastpage
4
Abstract
A new extension of the AD5933 impedance measurement chip is presented. It allows to use the chip in the four electrode configuration to reduce the artifact effects of interface impedance between the electrode and the sample to be measured. The measurement approach is validated on various test circuits, and the system has been able to measure small physiological samples (50 ul) in a four microelectrode configuration and with a low sensitivity to interface impedances. The system is a good candidate for any embedded bioimpedance measurement system which requires a low sensitivity to interface impedances.
Keywords
bioelectric phenomena; biomedical electrodes; biomedical electronics; biomedical measurement; electric impedance measurement; microelectrodes; AD5933 impedance measurement chip; bioimpedance measurement; four-electrode configuration; interface impedance; Current measurement; Electrodes; Frequency measurement; Impedance; Impedance measurement; Semiconductor device measurement; Voltage measurement;