    Four electrode embedded bioimpedance measurement system

  • Author

    Margo, Cedric ; Katrib, Juliano ; Nadi, M. ; Rouane, Amar

  • Author_Institution
    LPEM, ESPCI-ParisTech, Paris, France
  • fYear
    2013
  • fDate
    20-21 June 2013
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A new extension of the AD5933 impedance measurement chip is presented. It allows to use the chip in the four electrode configuration to reduce the artifact effects of interface impedance between the electrode and the sample to be measured. The measurement approach is validated on various test circuits, and the system has been able to measure small physiological samples (50 ul) in a four microelectrode configuration and with a low sensitivity to interface impedances. The system is a good candidate for any embedded bioimpedance measurement system which requires a low sensitivity to interface impedances.
  • Keywords
    bioelectric phenomena; biomedical electrodes; biomedical electronics; biomedical measurement; electric impedance measurement; microelectrodes; AD5933 impedance measurement chip; bioimpedance measurement; four-electrode configuration; interface impedance; Current measurement; Electrodes; Frequency measurement; Impedance; Impedance measurement; Semiconductor device measurement; Voltage measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Faible Tension Faible Consommation (FTFC), 2013 IEEE
  • Conference_Location
    Paris
  • Print_ISBN
    978-1-4673-6105-7
  • Type

    conf

  • DOI
    10.1109/FTFC.2013.6577773