Abstract :
A cavity ionization chamber, made of PVC coated with graphite, was developed and tested to verify its applicability as a reference dosimeter for 60Co beams. In order to characterize this new ionization chamber, several tests were made, following international recommendations, as saturation, ion collection efficiency, polarity effect, short- and medium-term stabilities, linearity of response, angular dependence and leakage current. All results obtained were within those recommended internationally, showing that this prototype may be used as a dosimetric standard system at the Calibration Laboratory of IPEN.
Keywords :
dosimetry; ionisation chambers; radiation therapy; 60Co beam dosimetry; IPEN Calibration Laboratory; cavity ionization chamber; cylindrical ionization chamber; dosimetric standard system; international recommendations; ion collection efήciency; leakage current; medium-term stability; polarity effect; radiation dosimetry; radiotherapy; short-term stability; Dosimetry; Ionization chambers; Laboratories; Materials; Standards; X-rays; Gamma beam dosimetry; ionization chamber; radiotherapy;