DocumentCode :
636454
Title :
Radiation dose and accuracy analysis of newly developed cone-beam CT for dental and maxillofacial imaging
Author :
Thongvigitmanee, Saowapak S. ; Pongnapang, Napapong ; Aootaphao, Sorapong ; Yampri, Pinyo ; Srivongsa, Tanapong ; Sirisalee, Pasu ; Rajruangrabin, Jartuwat ; Thajchayapong, P.
Author_Institution :
Image Technol. Lab., Nat. Sci. & Technol. Dev. Agency, Pathumthani, Thailand
fYear :
2013
fDate :
3-7 July 2013
Firstpage :
2356
Lastpage :
2359
Abstract :
Cone-beam computed tomography (CBCT) has become increasingly popular in dental and maxillofacial imaging due to its accurate 3D information, minimal radiation dose, and low machine cost. In this paper, we have proposed the newly developed CBCT scanner, called DentiiScan. Our gantry system consisting of a cone-beam X-ray source and an amorphous silicon flat panel detector is rotated around a patient´s head. With the large area detector, only a single rotation is needed to reconstruct the field-of-view area from chin to eyes and our reconstructed algorithm based on GPU calculation is about 30 times faster than the CPU-based algorithm. The radiation dose was measured and compared to other dental and medical CT machines. The absorbed radiation dose from our proposed CBCT machine is significantly low. In addition, geometric accuracy was analyzed when the test object was scanned at the normal position as well as the inclined position. The results from three observers repeated for five times confirm that the machine can produce reconstructed images with high accuracy.
Keywords :
computerised tomography; dosimetry; CBCT machine; CPU based algorithm; DentiiScan; GPU calculation; cone beam CT; cone beam computed tomography; dental imaging; field-of-view area reconstruction; gantry system; maxillofacial imaging; radiation accuracy analysis; radiation dose; Accuracy; Computed tomography; Dentistry; Image reconstruction; Medical diagnostic imaging; Three-dimensional displays;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location :
Osaka
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2013.6610011
Filename :
6610011
Link To Document :
بازگشت