DocumentCode :
651435
Title :
Low-power low-noise ECG acquisition system with dsp for heart disease identification
Author :
Bo-Yu Shiu ; Shuo-Wei Wang ; Yuan-Sun Chu ; Tsung-Heng Tsai
Author_Institution :
Dept. of Electr. Eng., Nat. Chung Cheng Univ., Chiayi, Taiwan
fYear :
2013
fDate :
Oct. 31 2013-Nov. 2 2013
Firstpage :
21
Lastpage :
24
Abstract :
In this paper, we implemented a low-power low-noise ECG acquisition system for heart disease identifications. The ECG signal is acquired with an analog front-end circuit, and the offset and baseline drift is eliminated at the same time. We also implemented a digital signal processing (DSP) unit to effectively remove EMG interferences. Interception of ST segment is proposed to achieve the identification of heart diseases. The ECG front-end chip has been designed and fabricated by using a TSMC 90nm CMOS technology, the total power consumption was measured at 40.3μW. DSP algorithms are carried out in the FPGA. Experimental results show that the sensitivity and specificity of ST segment classification after EMG elimination is 96.6% and 93.1%, respectively.
Keywords :
CMOS integrated circuits; bioelectric potentials; biomedical electronics; diseases; electrocardiography; medical signal processing; power consumption; signal classification; DSP unit; ECG front-end chip; ECG signal; EMG elimination; FPGA; ST segment classification; ST segment interception; TSMC CMOS technology; analog front-end circuit; baseline drift offset; digital signal processing unit; heart disease identification; heart diseases; low-power low-noise ECG acquisition system; power 40.3 muW; size 90 nm; total power consumption;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Circuits and Systems Conference (BioCAS), 2013 IEEE
Conference_Location :
Rotterdam
Type :
conf
DOI :
10.1109/BioCAS.2013.6679630
Filename :
6679630
Link To Document :
بازگشت