• DocumentCode
    655716
  • Title

    RF transmit performance comparison for several MRI head array geometries

  • Author

    Kozlov, M. ; Turner, Richard

  • Author_Institution
    Max Planck Inst. for Human Cognitive & Brain Sci., Leipzig, Germany
  • fYear
    2013
  • fDate
    6-10 Oct. 2013
  • Firstpage
    601
  • Lastpage
    604
  • Abstract
    We numerically investigated several magnetic resonance imaging radiofrequency transmit coil arrays with different array diameters and tuning conditions. Compared with a dual-row array of 230 mm in diameter, it is difficult to achieve high transmit performance when the diameter is increased to 280 mm, while maintaining high load independence and a predictable distribution of current through each element. All arrays investigated provided similar inhomogeneity when each array row was accurately excited in a circular polarization mode, and the phase shift between rows and asymmetry in the excitation power for each row were properly adjusted.
  • Keywords
    biomedical MRI; coils; computational electromagnetics; microwave antenna arrays; MRI head array geometries; RF transmit performance comparison; array diameters; circular polarization mode; dual-row array; excitation power; high transmit performance; magnetic resonance imaging radiofrequency transmit coil arrays; phase shift; size 230 nm; size 280 nm; tuning conditions; Arrays; Capacitors; Coils; Load modeling; Magnetic resonance imaging; Optimization; Tuning; Antenna arrays; Biomedical imaging; Computational electromagnetics;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference (EuMC), 2013 European
  • Conference_Location
    Nuremberg
  • Type

    conf

  • Filename
    6686727