Title :
Sensitivity and response time of an ozone sensor
Author :
David, Marcelo ; Marcus, Tay Ching En ; Yaacob, M. ; Salim, Mohd Rashidi ; Ibrahim, Mohd Haniff ; Idrus, S.M. ; IzamAzmi, Asrul ; Ngajikin, Nor Hafizah
Author_Institution :
Lightwave Commun. Res. Group, Univ. Teknol. Malaysia, Skudai, Malaysia
Abstract :
The use of optical retro-reflectors in improving the sensitivity and response time of an optical sensor based on optical absorption spectroscopy for the measurement of ozone gas is presented. Two optical retro-reflectors are employed in the design of a 30cm and 20 cm absorption gas cells. Our analysis shows that, in the 30cm gas cell, a sensitivity of 0.0451ppm and 0.0901ppm is achievable while in the 20cm gas cell we can achieve a sensitivity value of 0.0681ppm. However these sensitivity values are dependent on the optical density of the sensor. In general gas cell with wider diameters has potentials for a faster response time.
Keywords :
gas sensors; optical elements; optical sensors; ozone; spectrochemical analysis; O3; gas cell; optical absorption spectroscopy; optical density; optical retroreflectors; ozone sensor; response time; sensitivity; Biomedical optical imaging; Optical fiber sensors; Optical fibers; Optical polarization; Optical reflection; Monitoring; Optical path length; Optical recto-reflectors; gas cell diameter; ozone; response time; sensitivity;
Conference_Titel :
Photonics (ICP), 2013 IEEE 4th International Conference on
Conference_Location :
Melaka
Print_ISBN :
978-1-4673-6073-9
DOI :
10.1109/ICP.2013.6687064