DocumentCode :
656510
Title :
Analysis of backflow within an external centrifugal blood pump for ventricular assist device
Author :
Rungsirigulnan, Chitsanupong ; Laohasurayodhin, Ravi ; Tuanthammaruk, Thisana ; Chusri, Yanee ; Diloksumpan, Paweena ; Naiyanetr, Phomphop
Author_Institution :
Dept. of Biomed. Eng., Mahidol Univ., Nakhonpathom, Thailand
fYear :
2013
fDate :
23-25 Oct. 2013
Firstpage :
1
Lastpage :
4
Abstract :
During left ventricular support by rotary blood pumps (RBPs), pump speed plays an important role in the amount of heart support. The reducing of pump speed seems important in order to prevent some adverse events in the heart valve. However, when pump speed was reduced, the backflow which is the flow that has the reverse direction to blood pumping direction may occur more easily. This is due to the amount of pressure at an outlet of pump. Head pressure and pump flow curve (HQ-curve) is the graph that shows the performance of speed, head pressure and flow rate in the blood pump. In this study, two impellers from MUAD-VAD02 were used; 90-degree-angle straight blade (90SB) and 130 degrees straight blade (130BSB). The backflow of two impellers were investigated at three pump speeds (1000, 1400 and 1800 rpm) using HQ-curve, negative flow rate vs. time, and pressure vs. time. The experimental study uses in vitro testing by simulating the natural circulatory system with RBPs in the mock circulation loop system. The back flow of 90SB had more negative flow than BSB at all three speeds. In conclusion, BSB- impeller was chosen over the 90SB-impeller because at the same head pressure and speed in HQ-curve, the BSB impeller has less negative flow.
Keywords :
biomedical equipment; haemodynamics; impellers; patient treatment; pumps; 130-degrees straight blade; 90-degree-angle straight blade; MUAD-VAD02; RBP; backflow analysis; blood pump flow rate; external centrifugal blood pump; head pressure; heart support; heart valve; impellers; left ventricular support; mock circulation loop system; natural circulatory system; pump flow curve; pump speed; rotary blood pumps; ventricular assist device; Blades; Blood; Fluids; Heart; Impellers; Pumps; Valves; Backflow; HQ-curve; Mock circulation Introduction (Heading 1); RBPs;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Biomedical Engineering International Conference (BMEiCON), 2013 6th
Conference_Location :
Amphur Muang
Print_ISBN :
978-1-4799-1466-1
Type :
conf
DOI :
10.1109/BMEiCon.2013.6687729
Filename :
6687729