Safety assessment of software-intensive medical devices: Introducing a safety quality model approach
Author :
Adler, Rasmus ; Kemmann, Soeren ; de Melo Carvalho Filho, Djalma ; Oliveira Neto, Jose Augusto
Author_Institution :
Dept.: Embedded Syst. Quality Assurance, Fraunhofer IESE, Kaiserslautern, Germany
fYear :
2013
fDate :
4-7 Nov. 2013
Firstpage :
217
Lastpage :
222
Abstract :
Argumentation-based safety assurance is a promising approach for the development of safe software-intensive medical devices. However, one challenge is safety assessment by an independent authority. This article presents an approach that enables argumentation-based safety development on the one hand, while providing means for assessing the product´s safety afterwards on the other hand. We combine a generic safety case with an engineering model, which results in specific quality questions for assessors and provides a generic argumentation structure for manufacturers.
Keywords :
biomedical equipment; medical computing; safety-critical software; software quality; argumentation structure; argumentation-based safety assurance; argumentation-based safety development; independent authority; safety assessment; safety quality model approach; software-intensive medical devices; Certification; Context; Hazards; Product safety; Software; Standards; medical software; safety assessment; safety case;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Software Reliability Engineering Workshops (ISSREW), 2013 IEEE International Symposium on
Conference_Location :
Pasadena, CA
Type :
conf
DOI :
10.1109/ISSREW.2013.6688901
Filename :
6688901
Link To Document :
بازگشت