• DocumentCode
    663079
  • Title

    Stability and selectivity of a chronic, multi-contact cuff electrode for sensory stimulation in a human amputee

  • Author

    Tan, Don F.D. ; Schiefer, M. ; Keith, Michael W. ; Anderson, Richard ; Tyler, D.J.

  • Author_Institution
    Case Western Reserve Univ., Cleveland, OH, USA
  • fYear
    2013
  • fDate
    6-8 Nov. 2013
  • Firstpage
    859
  • Lastpage
    862
  • Abstract
    Multi-contact cuff electrodes were implanted in the median, ulnar, and radial nerves of an upper-limb amputee. 19 of 20 contacts produced selective, sensory response in the phantom limb from nerve stimulation. The neural interface has been stable for the duration of this ongoing chronic study, 12 months post-implant, with consistent threshold and impedance measures.
  • Keywords
    bioelectric potentials; biomedical electrodes; electric impedance measurement; neurophysiology; phantoms; prosthetics; chronic multicontact cuff electrode selectivity; chronic multicontact cuff electrode stability; human amputee; impedance measurement; median nerves; nerve stimulation; neural interface; phantom limb; post-implant study; radial nerves; selective sensory response; sensory stimulation; time 12 month; ulnar nerves; upper-limb amputee; Electrodes; Electronic mail; Impedance; Impedance measurement; Implants; Phantoms; Spirals;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Neural Engineering (NER), 2013 6th International IEEE/EMBS Conference on
  • Conference_Location
    San Diego, CA
  • ISSN
    1948-3546
  • Type

    conf

  • DOI
    10.1109/NER.2013.6696070
  • Filename
    6696070