DocumentCode :
674500
Title :
Design and testing of a cardiac monitor for home care
Author :
Lorenzo-Costa, Yaniesis ; Guardarrama-Mieres, Ronny ; Gonzalez-Fernandez, Rene
Author_Institution :
Central Inst. of Digital Res., Havana, Cuba
fYear :
2013
fDate :
22-25 Sept. 2013
Firstpage :
495
Lastpage :
498
Abstract :
The aim of this paper is to discuss the design and testing of a new battery-powered cardiac monitor for Homecare. The device is able to acquire two ECG channels; these signals can be processed, stored and transmitted through a Bluetooth´s communication channel. The monitor is based on the MSP430F5438 microcontroller from Texas Instruments and includes a graphic LCD, a simple keypad, a two-channel ECG amplifier, and two NiMH batteries. The ECG is displayed in real time with the heart rate; QRS complexes are detected to compute heart rate. The QRS complex detection is based on an energy function. A new element was introduced in the ECG acquisition; an ECG channel can be acquired by holding with hands the device because its case has two embedded electrodes. On the other hand, the two ECG channels can be acquired by connecting the appropriated patient cable to the monitor. The method developed for QRS complex detection has been tested with twelve ECG strips from MIT-BIH annotated ECG database; 98.05% of the QRS complexes were detected and false positives were not present.
Keywords :
Bluetooth; amplifiers; bioelectric potentials; biomedical electrodes; data communication; electrocardiography; health care; keyboards; liquid crystal displays; medical signal detection; medical signal processing; microcontrollers; telemedicine; Bluetooth communication channel; ECG channel acquisition; ECG signal processing; ECG signal storage; ECG signal transmission; MIT-BIH annotated ECG database; MSP430F5438 microcontroller; QRS complex detection; battery-powered cardiac monitor; electrocardiography; electrodes; energy function; graphic LCD; heart rate computation; home care; keypad; two-channel ECG amplifier; Abstracts; Electrocardiography; Electrodes; Heart; Real-time systems; Rhythm; Telemedicine;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computing in Cardiology Conference (CinC), 2013
Conference_Location :
Zaragoza
ISSN :
2325-8861
Print_ISBN :
978-1-4799-0884-4
Type :
conf
Filename :
6713422
Link To Document :
بازگشت