Title :
Evaluation of imaging modules based on SensL array SB-8 for nuclear medicine applications
Author :
Stolin, Alexander V. ; Majewski, S. ; Jaliparthi, Gangadhar ; Raylman, Raymond R. ; Proffitt, James
Author_Institution :
Dept. of Radiol., West Virginia Univ., Morgantown, WV, USA
fDate :
Oct. 27 2013-Nov. 2 2013
Abstract :
Recent advances in the development of silicon photomultipliers (SiPM) offer new opportunities for medical imaging applications. Specifically, novel imaging devices for positron emission (PET) and single photoemission computed tomography (SPECT) are becoming feasible. In this investigation, we tested a monolithic array of new generation SiPMs, an SB-8 array from SensL. 8×8 array of 6-mm square SiPMs was studied with two different multiplexing readouts, 4- and 16-channels. SB-8 detector was coupled to various scintillation arrays and resulting prototypes were evaluated. Testing demonstrated that the new device is capable of resolving 1 mm LYSO and 1.5 mm NaI crystals with 16-channel readout and 1.57 mm LYSO and 3 mm NaI crystals with 4-channel readout. Energy resolution of approximately 15 % at 511 keV and 19 % at 122 keV were obtained with LYSO and NaI crystals respectively. Timing resolution of 1.52 ns was measured with 1.57 mm LYSO array and the 16-channel readout. It is concluded that new B-series SiPMs from SensL are suitable for use in high performance nuclear medicine particle detectors.
Keywords :
biomedical equipment; elemental semiconductors; multiplexing; particle detectors; photomultipliers; positron emission tomography; scintillation; silicon; single photon emission computed tomography; B-series SiPMs; LYSO crystals; NaI crystals; PET; SB-8 detector; SPECT; SensL array SB-8 array; SensL. 8x8 array; Si; high performance nuclear medicine particle detectors; imaging devices; imaging modules; medical imaging applications; monolithic array; multiplexing readouts; nuclear medicine applications; positron emission tomography; scintillation arrays; silicon photomultipliers; single photoemission computed tomography; size 1 mm to 6 mm; timing resolution; Arrays; Crystals; Detectors; Energy resolution; Image resolution; Single photon emission computed tomography; Temperature measurement; Nuclear Medicine Imaging; PET; SPECT Instrumentation; SiPM;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location :
Seoul
Print_ISBN :
978-1-4799-0533-1
DOI :
10.1109/NSSMIC.2013.6829079