  • Abstract
    A method for reducing lost events in LOR rebinning using axially extended sinograms was evaluated in phantom and awake rat studies. The results suggest that a relatively small number of additional virtual detector rings leads to a sizeable reduction in lost events, reducing noise levels, and less affected by the blurring associated with the more simplistic SSRB approach. Although use of additional virtual rings increases computation load by requiring larger sinograms, we postulate that this may still provide a suitable `middle ground´ approach between SSRB and a full list mode reconstruction.
  • Keywords
    image denoising; image reconstruction; medical image processing; motion compensation; phantoms; positron emission tomography; SSRB approach; awake rat studies; axial field-of-view; event losses reduction; image blurring; image reconstruction; noise level reduction; phantom; positron emission tomography; sinogram-based PET motion correction; virtual detector rings; Animals; Approximation methods; Detectors; Image reconstruction; Phantoms; Positron emission tomography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
  • Conference_Location
    Seoul
  • Print_ISBN
    978-1-4799-0533-1
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2013.6829250
  • Filename
    6829250