Title :
Results of tests of three-dimensionally integrated chips bonded to sensors
Author :
Deptuch, Grzegorz W. ; Carini, Gabriella ; Collier, Terence ; Grybos, Pawel ; Kmon, Piotr ; Lipton, Ronald ; Maj, Piotr ; Trimpl, Marcel ; Siddons, David P. ; Szczygiel, Robert ; Yarema, Raymond
Author_Institution :
Electr. Eng. Dept., Fermi Nat. Accel. Lab., Batavia, IL, USA
fDate :
Oct. 27 2013-Nov. 2 2013
Abstract :
The VIPIC1 pixel readout integrated circuit was designed for high timing resolution, pixel based, X-ray Photon Correlation Spectroscopy experiments, typically using 8 keV X-rays at a synchrotron radiation facility. Technical hurdles related to a very aggressive alignment resulted in delays with delivery of 3D bonded wafers that eventually yielded operational devices. Recent improvements in alignment of wafers bonded with the Cu-Cu thermo-compression method and resorting to the Cu-DBI® bonding method started providing operational devices that were tested and led to the next steps. These were the tests with radioactive sources, emitting X-rays, executed after bonding of individual 3D dies to Silicon pixelated sensors for the first time. Results of these tests are presented.
Keywords :
X-ray spectroscopy; nuclear electronics; radioactive sources; readout electronics; silicon radiation detectors; wafer bonding; 3D bonded wafers; Cu-Cu thermocompression method; Cu-DBI® bonding method; VIPIC1 pixel readout integrated circuit; X-ray emission; X-ray photon correlation spectroscopy experiments; radioactive sources; silicon pixelated sensors; synchrotron radiation facility; three-dimensionally integrated chips; timing resolution; Adhesives; Nickel; Pediatrics; Photonics; Sensors; Three-dimensional displays;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2013 IEEE
Conference_Location :
Seoul
Print_ISBN :
978-1-4799-0533-1
DOI :
10.1109/NSSMIC.2013.6829434