Abstract :
We have evaluated configurations of 1-meter long ZnS/6LiF scintillator screens, wavelength shifting readout, and moderator to find an optimum meeting minimum standards for neutron portal monitoring applications. MCNP5 modeling predicts that by embedding 3 polyethylene cylinders covered with ZnS/6LiF scintillator in a rectangular moderator, a sensitivity of 2.6 cps/ng 252Cf at 2 m can be achieved.
Keywords :
californium; lithium compounds; neutron detection; nuclear electronics; readout electronics; solid scintillation detectors; zinc compounds; 252Cf sensitivity; 6LiF; MCNP5 modeling; ZnS; ZnS-6LiF moderated neutron detector design; embedding 3 polyethylene cylinders; long ZnS-6LiF scintillator screen configurations; neutron portal monitoring applications; optimum meeting minimum standards; rectangular moderator; wavelength shifting readout; Conferences; Detectors; Electron tubes; Geometry; Neutrons; Optical fiber devices; Polyethylene;