A New Algorithm of Drop Speed Measuring of Infusion Monitoring Device
Author :
Jiang Ma ; Yuqiao Wen ; Jie Yin ; Duoli Zhong ; Kun Yang ; Wei Zeng ; Lingyan Du ; Kang Huang
Author_Institution :
Dept. of Electron. Eng., Sichuan Univ. of Sci. & Eng., Zigong, China
fYear :
2013
fDate :
14-15 Dec. 2013
Firstpage :
229
Lastpage :
232
Abstract :
Alarm accuracy of infusion monitoring device mostly depends on drop speed measuring of infusion. A new drop speed calculation is put forward for the discontinuous characteristics of infusion. 5 drops is taken as a calculation unit. When new drop is detected, the time interval between new drop & its previous one will replace that between first one drop and second drop. The average drop speed will be calculated like that. Compared with the drop speed mentioned in Reference, this advantage of algorithms is of real time, reliability and effectively reducing wrong-alarm rate.
Keywords :
biomedical equipment; biomedical measurement; drops; reliability; alarm accuracy; discontinuous characteristics; drop speed measurement; infusion monitoring device; reliability; time interval; Current measurement; Educational institutions; Hospitals; Monitoring; Real-time systems; Safety; Time measurement; algorithms; drop speed; infusion monitoring; pushing forward-type;