Abstract :
The sensitivity of a PM FFL MPI (permanent magnet field-free line magnetic particle imaging) scanner was investigated using a serial dilution. The linearity of the reconstructed signal with respect to the particle concentration has been verified. A lower detection limit of 15 μg iron in magnetite has been found.
Keywords :
biomagnetism; biomedical MRI; image reconstruction; iron compounds; magnetic particles; medical image processing; permanent magnets; Fe3O4; MPI FFL scanner; magnetite; mass 15 mug; particle concentration; permanent magnet field-free line magnetic particle imaging; reconstructed signal linearity; sensitivity; serial dilution; Image reconstruction; Iron; Linearity; Magnetic particles; Magnetic resonance imaging; Sensitivity;