Title :
Characterization and correction of calibration jigs for LISN impedance measurements
Author_Institution :
Seibersdorf Labs., EMC & Opt., Seibersdorf, Austria
Abstract :
The papers deals with the influence of the calibration jig to the impedance measurement result. The goal is to characterize two different designs of calibration jigs and to correct the LISN impedance accordingly. Without proper correction there is a large difference of the phase angle. It is found that the correction works properly. However the results are not without ambiguity, due to the lack of the definition of a measurand. Assumptions regarding a suitable calibration plane must be made for proper phase angle measurements.
Keywords :
calibration; electric impedance measurement; electromagnetic interference; LISN impedance measurements; calibration jigs; line impedance network stabilization; phase angle difference; phase angle measurement; Calibration; Electromagnetic compatibility; Fixtures; Footwear; Impedance; Impedance measurement; Measurement uncertainty; calibration; equivalent circuit; impedance measurement; parasitic capacitance; phase measurement;
Conference_Titel :
Electromagnetic Compatibility and Signal Integrity, 2015 IEEE Symposium on
Conference_Location :
Santa Clara, CA
Print_ISBN :
978-1-4799-1992-5
DOI :
10.1109/EMCSI.2015.7107688