DocumentCode :
71548
Title :
Fast T2 Mapping With Improved Accuracy Using Undersampled Spin-Echo MRI and Model-Based Reconstructions With a Generating Function
Author :
Sumpf, Tilman J. ; Petrovic, Andreas ; Uecker, Martin ; Knoll, Florian ; Frahm, Jens
Author_Institution :
Max-Planck-Inst. for Biophys. Chem., Biomed. NMR Forschungs GmbH, Gottingen, Germany
Volume :
33
Issue :
12
fYear :
2014
fDate :
Dec. 2014
Firstpage :
2213
Lastpage :
2222
Abstract :
A model-based reconstruction technique for accelerated T2 mapping with improved accuracy is proposed using undersampled Cartesian spin-echo magnetic resonance imaging (MRI) data. The technique employs an advanced signal model for T2 relaxation that accounts for contributions from indirect echoes in a train of multiple spin echoes. An iterative solution of the nonlinear inverse reconstruction problem directly estimates spin-density and T2 maps from undersampled raw data. The algorithm is validated for simulated data as well as phantom and human brain MRI at 3T. The performance of the advanced model is compared to conventional pixel-based fitting of echo-time images from fully sampled data. The proposed method yields more accurate T2 values than the mono-exponential model and allows for retrospective undersampling factors of at least 6. Although limitations are observed for very long T2 relaxation times, respective reconstruction problems may be overcome by a gradient dampening approach. The analytical gradient of the utilized cost function is included as Appendix . The source code is made available to the community.
Keywords :
biomedical MRI; brain; gradient methods; image reconstruction; image sampling; inverse problems; medical image processing; phantoms; physiological models; spin echo (NMR); Appendix; T2 relaxation times; T2 values; accelerated T2 mapping; advanced model performance; advanced signal model; analytical gradient; conventional pixel-based fitting; cost function; echo-time images; fast T2 mapping; fully sampled data; generating function; gradient dampening approach; human brain MRI; improved accuracy; indirect echoes; iterative solution; magnetic flux density 3 T; model-based reconstruction technique; model-based reconstructions; mono-exponential model; multiple spin echo train; nonlinear inverse reconstruction problem; phantom; respective reconstruction problems; retrospective undersampling factors; simulated data; source code; spin-density; undersampled Cartesian spin-echo magnetic resonance imaging data; undersampled raw data; undersampled spin-echo MRI; Accuracy; Data models; Discrete Fourier transforms; Frequency-domain analysis; Image reconstruction; Magnetic resonance imaging; Phantoms; Fast spin echo (FSE); T2 mapping; indirect echoes; model-based reconstruction; relaxometry;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.2014.2333370
Filename :
6844895
Link To Document :
بازگشت