Title :
Factors affecting the performance of a SQUID gradiometer
Author :
Baltag, Octavian ; Rau, Miuta Carmina
Author_Institution :
Terraflux Control Ltd., Iasi, Romania
Abstract :
The work presents the errors induced by some geometrical imperfections underlying in the experimental realization of a 1st and 2nd order SQUID gradiometer with electronic subtraction. It will be discus the effect of the non identity coils of the magnetometer channels and the signal noise ratio for 1st and 2nd order magnetic field gradients for two different magnetic sources, one of disturbance and one of biosignal separated by a certain distance.
Keywords :
SQUID magnetometers; biomagnetism; SQUID gradiometer; biosignal; disturbance; electronic subtraction; geometrical imperfections; magnetic field gradients; magnetic sources; magnetometer channels; non identity coils; signal noise ratio; Coils; Magnetic field measurement; Magnetic fields; Measurement uncertainty; Noise; SQUIDs; Superconducting magnets; SQUID; biomagnetometry; gradiometer;
Conference_Titel :
Advanced Topics in Electrical Engineering (ATEE), 2015 9th International Symposium on
Conference_Location :
Bucharest
DOI :
10.1109/ATEE.2015.7133945