DocumentCode :
717376
Title :
A 7T double-tuned (1H/31P) microstrip surface RF coil for the IMAGO7 MR scanner
Author :
Retico, Alessandra ; Vitacolonna, Assunta ; Galante, Angelo ; Florio, Tiziana M. ; Cimini, Annamaria ; Stara, Riccardo ; Tiberi, Gianluigi ; Tosetti, Michela ; Fontana, Nunzia ; Manara, Giuliano ; Monorchio, Agostino ; Alecci, Marcello
Author_Institution :
Sez. Pisa, Ist. Naz. di Fis. Nucl. (INFN), Pisa, Italy
fYear :
2015
fDate :
7-9 May 2015
Firstpage :
84
Lastpage :
88
Abstract :
Ultra-High Field (UHF) (4-9.4T) Magnetic Resonance Imaging (MRI) and Spectroscopy (MRS) are valuable tools in the diagnosis and monitoring of many diseases thanks to the enhanced Signal-to-Noise Ratio (SNR) and spectral/spatial resolution. However, such UHF MR applications require the development and optimization of specially designed Radio Frequency (RF) coils. In this study we report the design, construction, workbench and MR testing of a novel 7T double-tuned (1H/31P) RF surface coil made with microstrip technology. The RF coil is suitable for MRI and MRS human studies of the lower limbs and also of the human head. The double-tuned RF coil was tested at the IMAGO7 Foundation in Pisa (Italy). Ex vivo MR 1H images and 31P spectra obtained with a large size fresh calf veal sample showed an excellent SNR within about 50 mm from the RF surface coil. It is believed that the combined 7T MR 1H/31P will provide benefits in performing clinical follow-up of muscular disorders in young and aging patients. The proposed design can be easily extended to the simultaneous detection of the 1H and other low-gamma nuclei, such as 7Li, 11B, 13C, 17O, 23Na.
Keywords :
biomedical MRI; coils; diseases; medical disorders; 7T MR 1H-31P; 7T double-tuned microstrip surface RF coil; IMAGO7 MR scanner; IMAGO7 foundation; MR 1H images; MR testing; UHF MR application; disease diagnosis; disease monitoring; human head; low-gamma nuclei; lower limb; microstrip technology; muscular disorder; radiofrequency coil; spectral-spatial resolution; ultrahigh field magnetic resonance imaging; ultrahigh field magnetic resonance spectroscopy; Magnetic resonance imaging; Microstrip; Radio frequency; Signal to noise ratio; Spectroscopy; Testing; 7T; Magnetic Resonance Imaging (MRI); Magnetic Resonance Spectroscopy (MRS); double-tuned; microstrip; radio-frequency coil;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Medical Measurements and Applications (MeMeA), 2015 IEEE International Symposium on
Conference_Location :
Turin
Type :
conf
DOI :
10.1109/MeMeA.2015.7145177
Filename :
7145177
Link To Document :
بازگشت