Title :
X-ray 4D stereoscopic technology
Author :
Miroshnychenko, S.I. ; Volkov, E.V.
Author_Institution :
“Teleoptic PRA” Ltd., Kiev, Ukraine
Abstract :
4D stereo X-ray imaging is realized as the sequence of 3D stereo images in some period of time. Data for 4D stereo can be received on tomosynthesis, linear tomography or cone beam tomography units. Majority of computed tomography systems don´t suits for 4D because of motion artifacts caused by organ´s deformation during data collecting. Organ´s movement causes geometrical measuring errors in 3D and 4D imaging. But such errors can be predicted and corrected. Correction of motion errors is based on predicted Z coordinate value for next stereo image and calculated value from acquired image.
Keywords :
X-ray imaging; computerised instrumentation; image motion analysis; image sequences; measurement errors; stereo image processing; 3D stereo image sequence; 4D stereo X-ray imaging; X-ray 4D stereoscopic technology; Z coordinate value prediction; cone beam tomography unit; data collection; geometrical measurement error; linear computed tomography system; motion artifact; motion error correction; organ deformation; tomosynthesis; Biomedical imaging; Computed tomography; Electron tubes; Stereo image processing; Three-dimensional displays; X-ray imaging; 3D; 4D; X-ray; anaglyph; digital receptor; geometrical measuring; stereo effect; tomosynthesis;
Conference_Titel :
Electronics and Nanotechnology (ELNANO), 2015 IEEE 35th International Conference on
Conference_Location :
Kiev
DOI :
10.1109/ELNANO.2015.7146905