DocumentCode :
718484
Title :
Dynamic errors of forced expiration measurements by spirometers
Author :
Lopata, Viktor ; Popov, Anton ; ElShabbah, M. ; Tomashevskiy, Roman ; Myasnyi, Ivan
Author_Institution :
Phys. & Biomed. Electron. Dept., Nat. Tech. Univ. of Ukraine, Kiev, Ukraine
fYear :
2015
fDate :
21-24 April 2015
Firstpage :
406
Lastpage :
408
Abstract :
Based on the forced expiration modeling, dynamic errors of process parameters measurements for spirometers with transducers of Lilly tube and turbine are calculated. It is shown that when the time constants of transducer are less than 3 ms, these errors for both types of spirometers do not exceed 0,5%.
Keywords :
biomedical equipment; measurement errors; pneumodynamics; transducers; turbines; Lilly tube; dynamic errors; forced expiration measurements; process parameters measurements; spirometers; time constants; transducers; turbine; Dynamics; Electron tubes; Mathematical model; Measurement errors; Measurement uncertainty; Transducers; Turbines; dynamic errors; forced expiration; spirometer;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electronics and Nanotechnology (ELNANO), 2015 IEEE 35th International Conference on
Conference_Location :
Kiev
Type :
conf
DOI :
10.1109/ELNANO.2015.7146920
Filename :
7146920
Link To Document :
بازگشت