Title :
Studies on calibration of IR radiometer by radiometric method
Author :
Kaur, S. ; Pandey, Sm ; Bhusare, Se
Author_Institution :
High Energy Mater. Res. Lab., Pune, India
Abstract :
IR radiometers are used for the measurement of IR intensity in military, industry, academic and research domains. For real-time, noncontact, accurate measurement with high sensitivity SR 5000 radiometer is employed. It covers the waveband from 0.4-14.5 μm and has a provision to accommodate many filters and detectors. The measurements are carried out with respect to internal black body, maintained at room temperature. As any measurement equipment requires repeatable, reproducible & accurate output, calibration at regular interval is a mandatory requirement. In the current paper the calibration of radiometer in radiometric mode using different filter bands at a distance of 80 and 3 meters is described. The calibration is done at different temperatures and Planck´s function is calculated. The measured results are validated with the Planck´s function. This paper also describes the convergence of results for 80 and 3 meters.
Keywords :
calibration; infrared detectors; intensity measurement; radiometers; radiometry; IR intensity measurement; IR radiometric method; Planck´s function; calibration; detector; distance 3 m; distance 80 m; filter band; high sensitivity SR 5000 radiometer; internal black body; temperature 293 K to 298 K; wavelength 0.4 mum to 14.5 mum; Artificial intelligence; Calibration; Convergence; Detectors; Monitoring; Radiometry; Wavelength measurement; BlackBody; Calibration; Detectors; Field of View; Filters; Radiometric Mode;
Conference_Titel :
Industrial Instrumentation and Control (ICIC), 2015 International Conference on
Conference_Location :
Pune
DOI :
10.1109/IIC.2015.7150721