• DocumentCode
    719966
  • Title

    Coupling analysis in the calibration process of electro-optical detection systems

  • Author

    Qijian Tang ; Xiangjun Wang ; Qingping Yang

  • Author_Institution
    MOEMS Educ. Minist. Key Lab., Tianjin Univ., Tianjin, China
  • fYear
    2015
  • fDate
    11-14 May 2015
  • Firstpage
    305
  • Lastpage
    308
  • Abstract
    Electro-optical detection systems should be calibrated prior to use, a commonly used method with an autocollimator and a high precision turntable may introduce large error in the calibration process. This paper analyzed the calibration process, and showed that the nature of the problem was attributed to the plane rotation around a non-orthogonal axis. The mirror´s non-orthogonal rotation was divided into position rotation and mirror spinning, and the effects on autocollimator readout were obtained. Simulation results demonstrated that the coupling effects on elevation and azimuth become serious as the inclined angle and rotated angle increase. It is therefore necessary to separate the coupling error in the calibration process, which would provide more accurate data for further compensation.
  • Keywords
    calibration; electro-optical devices; mirrors; optical collimators; photodetectors; calibration process; coupling analysis; electro-optical detection systems; mirror nonorthogonal rotation; mirror spinning; optical autocollimator readout; plane rotation; position rotation; Azimuth; Calibration; Couplings; Mathematical model; Mirrors; Quaternions; Spinning; EODS; calibration process; coupling analysis; quaternion; rotation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference (I2MTC), 2015 IEEE International
  • Conference_Location
    Pisa
  • Type

    conf

  • DOI
    10.1109/I2MTC.2015.7151284
  • Filename
    7151284